Call 08048601717 73% Response Rate

our range

Anaerobic Workstations

Leading Manufacturer of ia 500 anaerobic workstation, ia 300 anaerobic workstation and ia 800 anaerobic workstation from Thane.
IA 500 Anaerobic Workstation

IA 500 Anaerobic Workstation

Price on Request

Product Brochure

MaterialOpaque Sheet Of Acrylic Resin
External Dimension1200 X 740 X 660 mm
Main Chamber Dimension900 X 500 X 560 mm
Inter lock Dimension250 X 300 X 300 mm
Main Workstation CapacityUp to 600 Plates
Plates Size90 mm X 15 mm
Inter lock CapacityUp To 40 Plates
Interlock Cycle Time35 sec
Display TypeTouch Screen LCD Display
Oxygen Controlled With Display0 % To 22 %
Temperature Uniformity+- 1 Degree C at 37 DegreeC
Operating Temperature25 DegreeC to 45 DegreeC
Storage Temperatureup to 45 DegreeC
Relative HumidityUp to 90 % No condensing
Voltage220 V-240 V AC
Frequency50 Hz
ModelIA 500
Thickness6 mm

Specifications of “IA 500 Anaerobic Workstation":

Body Construction: Durable, Non-Corrosive, Opaque sheet of Acrylic Resin of 6mm or 8mm thickness for entire body, except front panel of acrylic resin but transparent sheet, for better visibility and insulation.

LCD display should continuously display and controlled oxygen in the chamber.

Oxygen should be controlled from 0.0 % to 22.0%.

Oxygen Controlled with Display 0 % to 22 % Starting from 0.2, 0.4,0.6, 0.8 upto 22%.

Workstation should be work as a Anaerobic workstation when Oxygen reach at 0.0%

To controlled the chamber oxygen, automatic forced purging system for Main Chamber as well as Interlock chamber, with adjustable purging period.

To achieve the chamber oxygen faster rate, additional facility to condition the workstation atmosphere using proper direction of internal air, should be provided.


LCD Display with touch screen Controlled Continuously show:

Chamber pressure

Humidity

Temperature

Fluorescent Light with hour meter

UV Light with hour meter

Internal Socket and Blower indication should be continuously present on the screen.

LCD display should continuously indicatePositive pressure (0.15 KPA/0.0015 bar) of the main chamber.

Audio & Visual alarm for low or excess oxygen level shall be provided.

Audio & Visual alarm for gas cylinder pressures, chamber pressure and temperature shall be provided.


Said Touch screen simplifies the calibration process: 

Automatically calibrates O2 sensor through one touch screen control.

Provides rapid equalization

Facilitates the download of data

Provides Faculty Recognition.


Workstation shall be supplied with Gas Control Software for:

USB communications port, 5 days continuous storage of event log data.

One touches calibration (O2), On-screen fault assistance, Cycle programming.


Safety Features:

Excess pressure relief valve

Temperature overshoot controller.

Condensate controller technology should be inbuilt in the workstation.

Pressure relief valve for releasing excess pressure inside chamber, shall be provided.

Heating element along with controller shall be provided to maintain the temperature inside chamber within ambient to 50°C, with Temperature Uniformity ± 1° C at 37°C.

Inbuilt Trim-point for temperature calibration shall be provided.

Workstation should be front 10% slope viewing panel.

There should be utility port with “flexible hose” for manual purging by Nitrogen, inside the chamber.

Workstation shall be operated by foot switch outside chamber.

Similar Foot Switch shall be provided for operation for gas nozzle ports & spot lamp inside chamber.

Catalyst and Activated Detox should be provided.


Standards & Safety :

CE Marked product, EN 12469

ISO 9001 : 2015

GMP Certified Product

Annual Calibration for Pressure, Temperature, Humidity, is included in scope.

View Complete details

get quoterequest a call back
  • IA 300 Anaerobic Workstation
  • IA 300 Anaerobic Workstation
IA 300 Anaerobic Workstation

IA 300 Anaerobic Workstation

Price on Request

Product Brochure

External Dimension900 X 700 X 575 mm
Internal Dimension650 X 475 X 475 mm
Interlock Dimension200 X 250 X 250 mm
Main Workstation CapacityUp to 350 Plates
Plate Size90 mm X 15 mm
Inter lock CapacityUp to 24 Plates
Interlock Cycle Time25 sec
Temperature Uniformity+- 1 Degree C at 37 DegreeC.
Operating Temperature25 DegreeC to 45 DegreeC
Storage Temperatureup to 45 DegreeC
Relative HumidityUp to 90 % No condensing
Voltage220 V-240 V AC
Frequency50 Hz
StandardEN 12469
ModelIA 300
Specifications of “IA 300 Anaerobic Workstation":
  • Body Construction: Durable, Non-Corrosive, Opaque sheet of Acrylic Resin of 6mm or 8mm thickness for entire body, except front panel of acrylic resin but transparent sheet, for better visibility and insulation.
  • LCD display should continuously display and controlled oxygen in the chamber.
  • Oxygen should be controlled from 0.0 % to 22.0%.
  • Oxygen Controlled with Display 0 % to 22 % Starting from 0.2, 0.4,0.6, 0.8 upto 22%.
  • Workstation should be work as a Anaerobic workstation when Oxygen reach at 0.0%
  • To controlled the chamber oxygen, automatic forced purging system for Main Chamber as well as Interlock chamber, with adjustable purging period.
  • To achieve the chamber oxygen faster rate, additional facility to condition the workstation atmosphere using proper direction of internal air, should be provided.

LCD Display with touch screen Controlled Continuously show:
  • Chamber pressure
  • Humidity
  • Temperature
  • Fluorescent Light with hour meter
  • UV Light with hour meter
  • Internal Socket and Blower indication should be continuously present on the screen.
  • LCD display should continuously indicatePositive pressure (0.15 KPA/0.0015 bar) of the main chamber.
  • Audio & Visual alarm for low or excess oxygen level shall be provided.
  • Audio & Visual alarm for gas cylinder pressures, chamber pressure and temperature shall be provided.

Said Touch screen simplifies the calibration process: 
  • Automatically calibrates O2 sensor through one touch screen control.
  • Provides rapid equalization
  • Facilitates the download of data
  • Provides Faculty Recognition.

Workstation shall be supplied with Gas Control Software for:
  • USB communications port, 5 days continuous storage of event log data.
  • One touches calibration (O2), On-screen fault assistance, Cycle programming.

Safety Features:
  • Excess pressure relief valve
  • Temperature overshoot controller.
  • Condensate controller technology should be inbuilt in the workstation.
  • Pressure relief valve for releasing excess pressure inside chamber, shall be provided.
  • Heating element along with controller shall be provided to maintain the temperature inside chamber within ambient to 50°C, with Temperature Uniformity ± 1° C at 37°C.
  • Inbuilt Trim-point for temperature calibration shall be provided.
  • Workstation should be front 10% slope viewing panel.
  • There should be utility port with “flexible hose” for manual purging by Nitrogen, inside the chamber.
  • Workstation shall be operated by foot switch outside chamber.
  • Similar Foot Switch shall be provided for operation for gas nozzle ports & spot lamp inside chamber.
  • Catalyst and Activated Detox should be provided.

Standards & Safety :
  • CE Marked product, EN 12469
  • ISO 9001 : 2015
  • GMP Certified Product
  • Annual Calibration for Pressure, Temperature, Humidity, is included in scope.

View Complete details

get quoterequest a call back
IA 800 Anaerobic Workstation

IA 800 Anaerobic Workstation

Price on Request

Product Brochure

MaterialOpaque Sheet Of Acrylic Resin
External Dimension1400 X 740 X 660 mm
Main Chamber Dimension1000 X 500 X 560 mm
Inter Lock Dimension300 X 350 X 300 mm
Main Workstation CapacityUp to 750 Plates
Plate Size90 mm X 15 mm
Inter Lock CapacityUp To 60 Plates
Interlock Cycle Time45 sec
Positive Pressure0.0015 bar
Oxygen Controlled With Display0 % to 22 %
Display TypeLCD Display
Relative HumidityUp to 90 % No condensing
Voltage220 V-240 V AC
Frequency50 Hz
ModelIA 800
Thickness8 mm

Specifications of “IA 800 Anaerobic Workstation":

Body Construction: Durable, Non-Corrosive, Opaque sheet of Acrylic Resin of 6mm or 8mm thickness for entire body, except front panel of acrylic resin but transparent sheet, for better visibility and insulation.

LCD display should continuously display and controlled oxygen in the chamber.

Oxygen should be controlled from 0.0 % to 22.0%.

Oxygen Controlled with Display 0 % to 22 % Starting from 0.2, 0.4,0.6, 0.8 upto 22%.

Workstation should be work as a Anaerobic workstation when Oxygen reach at 0.0%

To controlled the chamber oxygen, automatic forced purging system for Main Chamber as well as Interlock chamber, with adjustable purging period.

To achieve the chamber oxygen faster rate, additional facility to condition the workstation atmosphere using proper direction of internal air, should be provided.


LCD Display with touch screen Controlled Continuously show:

Chamber pressure

Humidity

Temperature

Fluorescent Light with hour meter

UV Light with hour meter

Internal Socket and Blower indication should be continuously present on the screen.

LCD display should continuously indicatePositive pressure (0.15 KPA/0.0015 bar) of the main chamber.

Audio & Visual alarm for low or excess oxygen level shall be provided.

Audio & Visual alarm for gas cylinder pressures, chamber pressure and temperature shall be provided.


Said Touch screen simplifies the calibration process: 

Automatically calibrates O2 sensor through one touch screen control.

Provides rapid equalization

Facilitates the download of data

Provides Faculty Recognition.


Workstation shall be supplied with Gas Control Software for:

USB communications port, 5 days continuous storage of event log data.

One touches calibration (O2), On-screen fault assistance, Cycle programming.


Safety Features:

Excess pressure relief valve

Temperature overshoot controller.

Condensate controller technology should be inbuilt in the workstation.

Pressure relief valve for releasing excess pressure inside chamber, shall be provided.

Heating element along with controller shall be provided to maintain the temperature inside chamber within ambient to 50°C, with Temperature Uniformity ± 1° C at 37°C.

Inbuilt Trim-point for temperature calibration shall be provided.

Workstation should be front 10% slope viewing panel.

There should be utility port with “flexible hose” for manual purging by Nitrogen, inside the chamber.

Workstation shall be operated by foot switch outside chamber.

Similar Foot Switch shall be provided for operation for gas nozzle ports & spot lamp inside chamber.

Catalyst and Activated Detox should be provided.


Standards & Safety :

CE Marked product, EN 12469

ISO 9001 : 2015

GMP Certified Product

Annual Calibration for Pressure, Temperature, Humidity, is included in scope.

View Complete details

get quoterequest a call back

EXPLORE MORE CATEGORIES